MIC Chinese Simplified Chinese Traditional English

Home » Product Directory » Electrical & Electronics » Electronic Instrument » Thermocouple Calibrator (V02)

Add to Basket

Thermocouple Calibrator (V02)

Thermocouple Calibrator (V02)
Thermocouple Calibrator (V02)

Product Description

Note:
1. Not including the accuracy of inner temperature compensatory transducer.
The range of inner temperature compensatory transducer: -10-50º C, compensatory error ¡Ü0.5º C.
2. Temperature coefficient: ± 0.005%, range/º C(0º C-18º C, 28º C-50ºC)

Input
 
Range
 
Input range
 
Resolution
 
Accuracy
 
Explanation
 
Resistance
 
400Ω
 
0.0~450.0Ω
 
0.1Ω
 
±0.05% Setting value ±0.2Ω
 
Measurong current 1mA Note1
 
Thermoresis-tance
 
Pt100
 
-200.0~850.0 º C
 
0.1 º C
 
±0.05% Setting value ±0.6 º C
 
1mA exciting current use Pt100-385
 
Measuring current 1mA Note1.2
 
Cu50
 
-50.0~150.0 º C
 
0.1 º C
 
1mA excitong current Note1.2
 
Input
 
Range
 
Input range
 
Resolution
 
Accuracy
 
Explanation
 
Resistance
 
400Ω
 
0.0~400.0Ω
 
0.1Ω
 
±0.05% Setting value ±0.2Ω
 
±1mA exciting current Note1.2
 
Thermoresis-tance
 
Pt100
 
-200.0~850.0 º C
 
0.1 º C
 
±0.05% Setting value ±0.6 º C
 
±1mA exciting current use Pt100-85
 
Note1.2
 
Cu50
 
-50.0~150.0 º C
 
0.1 º C
 
±1mA excitong current Note1.2
 
Power
 
9V battery (ANSI/NEDA 1640A or IEC 6LR619 alkaline) or AC power adapter
 
Battery
 
Approx. 20 hours under the condition of 10mA
 
Max.allowable voltage
 
30V
 
Operation temperture range
 
0-50°C
 
Operation humidity range
 
≤80%RH
 
Store temperature range
 
≤-10°C-50°C
 
Store humidity range
 
≤90%RH
 
Size
 
200×100×40mm
 
Weight
 
550g(with holster)
 
Accessories
 
Operation manual,test line CF-36(probe with alligator clip)
 
Options
 
AC adapter(VCPS),test line CF-31-A(probe clip)
 
Security
 
Conforms to IEC 1010
 
Model: V02
Company: Wenzhou Shangtong Instruments Co., Ltd.

Click here to send inquiry

Most Popular